Volume 102, Issue 11, November 2008, Pages 1667-1671
Respiratory Medicine

Short communication
Comparison of exhaled nitric oxide measurements between NIOX MINO® electrochemical and Ecomedics chemiluminescence analyzer

https://doi.org/10.1016/j.rmed.2008.06.021Get rights and content
Under an Elsevier user license
open archive

Summary

Background

Exhaled nitric oxide (eNO) is an established, noninvasive biomarker of active airway inflammation in (atopic) asthma. Treatment with anti-inflammatory therapy, such as inhaled corticosteroids, effectively decreases eNO levels. The NIOX MINO® (MINO) is a hand-held, relatively inexpensive, electrochemical device that has been shown to yield comparable eNO measurements to the NIOX stationary unit.

Aim

To compare measurements of MINO with another widely used and validated stationary chemiluminescence analyzer, the Ecomedics (ECO).

Methods

We performed subsequent eNO measurements on ECO and MINO in 50 subjects (19 healthy volunteers, 18 healthy smokers and 13 non-smoking, atopic asthmatics, not on controller therapy) on two visits 4–10 days apart. The mean of three acceptable measurements by ECO and the first acceptable measurement with the MINO were used for analysis.

Results

Both devices yielded reproducible eNO values for all subjects on both visits, with an overall CV of 22.7% (ECO) and 18.3% (MINO). A significant correlation was found between both devices (r = 0.97, p < 0.0001). Bland-Altman plots showed a high degree of agreement for the entire study population (mean difference MINO vs ECO = −10%; 95% limit of agreement were −36% and +28%) and in the three individual subgroups.

Conclusions

Exhaled NO values measured with the MINO are reproducible and in agreement with the ECO. Our results add further evidence to the reliability of the MINO and warrant its applicability in research and clinical practice.

Keywords

Airway inflammation
Exhaled NO
NIOX MINO®
Ecomedics
Reproducibility