Device | Correct technique | Errors in technique |
“Press and breathe” pMDI | Remove mouthpiece cap | Failure to remove mouthpiece cap* |
Shake inhaler (suspensions only) | Inhaler not shaken | |
Hold inhaler upright | Inhaler upside down* | |
Breathe out | No exhalation | |
Place mouthpiece between lips | ||
Fire while breathing in deeply and slowly | Firing device before start of inhalation | |
Firing device at or after end of inhalation* | ||
Inhaling through nose* | ||
Continue to inhale after firing | Stopping inhalation as device is fired* | |
Fast inhalation | ||
Hold breath (10 s) | No/short breath-hold | |
Breath-actuated pMDI | Remove mouthpiece cap | Failure to remove mouthpiece cap* |
Shake inhaler (suspensions only) | Inhaler not shaken | |
Hold inhaler upright | Inhaler upside down* | |
Prepare device (e.g. lift lever) | Failure to prepare device correctly* | |
Breathe out | No exhalation | |
Place mouthpiece between lips | Poor seal around mouthpiece | |
Using “open mouth” inhalation technique* | ||
Breathe in deeply and slowly | Weak inhalation, failure to trigger device* | |
Inhaling through nose* | ||
Continue to inhale after firing | Stopping inhalation as device is fired* | |
Fast inhalation | ||
Hold breath (10 s) | No/short breath-hold | |
“Press and breathe” pMDI plus spacer | Inappropriate handling (static charge) | |
Remove mouthpiece cap | Failure to remove mouthpiece cap* | |
Shake inhaler (suspensions only) | Inhaler not shaken | |
Hold inhaler upright | Inhaler upside down* | |
Insert pMDI into spacer | ||
Breathe out | No exhalation | |
Fire while breathing in deeply and slowly | Long delay before inhalation | |
Multiple actuation | ||
Weak inhalation, failure to open valve* | ||
Inhaling through nose* | ||
Continue to inhale after firing | Stopping inhalation as device is fired* | |
Fast inhalation | ||
Hold breath (10 s) | No/short breath-hold | |
DPIs | Remove cover (device specific) | Failure to remove cover* |
Load dose (device specific) | Incorrect dose loading* | |
Pierce capsule (single-dose devices) | Failure to pierce capsule* | |
Breathe out | Breathing out into device* | |
Place mouthpiece between lips | Inhalation vents blocked¶ | |
Poor seal round mouthpiece | ||
Using “open-mouth” inhalation technique* | ||
Inhale deeply and quickly | Not inhaling quickly enough* | |
Insufficient “acceleration” | ||
Inhaling through nose* | ||
Hold breath (10 s) | No/short breath-hold | |
Store in cool dry place | Inappropriate storage | |
Respimat® Soft MistTM Inhaler# | Hold upright and turn base | Failure to prime device/load dose* |
Open mouthpiece cap | Failure to open mouthpiece cap* | |
Breathe out | No exhalation | |
Put mouthpiece between lips | Mouthpiece vents blocked¶ | |
Press dose release button while breathing in deeply and slowly | Firing device before start of inhalation | |
Firing device at or after end of inhalation* | ||
Inhaling through nose* | ||
Continue to inhale after firing | Stopping inhalation as device is fired* | |
Fast inhalation | ||
Hold breath (10 s) | No/short breath-hold |
pMDI: pressurised metered-dose inhaler; DPI: dry powder inhaler. *: crucial error, likely to result in zero lung deposition of drug; #: manufactured by Boehringer Ingelheim GmbH & Co. KG, Ingelheim, Germany; ¶: error that may be crucial.