© 2005 the European Respiratory Society
Inhaler treatment options in COPDCORRESPONDENCE: S. P. Newman, Scientific Consultant, Nottingham, UK. E-mail: steve.newman{at}physics.org
A variety of inhaler devices are available for delivering treatments to patients with chronic obstructive pulmonary disease, and new inhalers are currently being developed. Each type of device has advantages and disadvantages, and the methods of preparation and use vary between them. The differences in instructions for use can easily confuse patients and health providers alike, resulting in incorrect use of many inhalers. "Crucial" errors in inhaler technique, whereby no drug is deposited in the lungs, must be avoided. Any type of inhaler can be misused so that little or no drug is deposited in the lungs. It is now increasingly widely recognised that a successful treatment outcome in chronic obstructive pulmonary disease depends as much on the inhaler device as it does on the drug. Inhaler choice in chronic obstructive pulmonary disease should take into account whether the patient is likely to use it correctly, as well as patient preference and the likelihood of adherence to treatment.
KEYWORDS: Chronic obstructive pulmonary disease, dry powder inhaler, pressurised metered-dose inhaler, soft mist inhaler, spacer device The inhaled route is preferred for the delivery of bronchodilators and corticosteroids used in the maintenance therapy of asthma and chronic obstructive pulmonary disease (COPD). Small doses of drugs are delivered direct to their site of action, leading to a rapid onset of action and a low incidence of side-effects. While guidance about appropriate selection of drugs for inhalation therapy is easily obtained [13], it is more difficult to source advice about the choice of inhaler device, despite the availability of several different types. The American College of Chest Physicians (ACCP) recently published evidence-based guidelines that listed eight points for consideration when selecting an inhaler, including taking account of any specific device preference the patient or clinician might have and whether a given patient can use the device properly [4]. Hence, an understanding of inhaler technology and issues of correct versus incorrect use are key factors affecting inhaler choice.
In this article, the main types of inhaler available for COPD therapy will be reviewed briefly, together with current understanding about correct and incorrect techniques for each device (table 1
It is assumed that inhaled drugs must be deposited within the lungs if they are to exert a beneficial effect [5]. Therefore, it is important to distinguish between "crucial" (or "essential") errors, which are likely to result in no drug reaching the lungs, and "non-crucial" errors, which are likely to result in a reduced amount of drug reaching the lungs compared with that attained using the correct technique [6]. Not all crucial errors are equally common in practice; table 3
The pMDI was first introduced in 1956 to provide a delivery system for inhaled bronchodilators with a multi-dose capability and reproducible dosing characteristics [8]. pMDIs contain propellants, which are currently being changed from chlorofluorocarbons (CFCs) to hydrofluoroalkanes (HFAs) because the former damage the ozone layer in the stratosphere [9]. The pMDI produces a rapid-moving plume of aerosol, the duration of which is typically 0.10.4 s [10]. The velocity of the aerosol plume may be 8 m·s1 at a distance of 10 cm from the actuator, and is even higher at distances closer to the nozzle [10]. The plume often feels cold on the back of the throat as the propellants evaporate. Most pMDIs only deposit 1020% of the dose in the lungs, even with good inhaler technique, and most of the dose is deposited in the oropharynx. Higher lung deposition and lower oropharyngeal deposition may be achieved with some recent formulations, where the drug is formulated as a solution in HFA propellant, rather than as a suspension of micronised particles [11]. Correct pMDI technique involves firing the pMDI, while breathing in deeply and slowly, and then following inhalation with a breath-holding pause to allow particles to sediment on the airway surfaces [12, 13]. Most instructions recommend placing the mouthpiece between closed lips, but it's also possible to hold the inhaler between open lips or even a few centimetres from the open mouth [14]. Misuse of pMDIs is common among patients [15, 16] and poor understanding of how to use a pMDI also appears to be widespread amongst health professionals [17, 18]. Most importantly, the pMDI must not be fired after the patient has completed inhalation, as there will then be no airstream to carry the aerosol into the lungs. Some aerosol will probably still reach the lungs if the pMDI is fired shortly before inhalation begins. Failure to correctly time firing with inhalation is sometimes termed "poor coordination" [19]. The second major problem with pMDI use is the so-called "cold Freon" effect, where the cold blast of propellants causes patients either to stop inhaling or to inhale via the nose [20]; this is probably less marked with HFA formulations. Patients who make errors in inhalation technique derive less clinical benefit than those with good technique [21], and this is particularly marked in those with poor coordination [22]. Despite the difficulties of using them correctly, pMDIs remain popular for delivering inhaled therapies in asthma and COPD because of their practical advantages: pMDIs contain at least 100 doses and are compact, portable, convenient and relatively inexpensive.
Since poor coordination between firing and inhaling is usually considered to be the most significant problem patients have with pMDIs [19], the development of breath-actuated (BA) pMDIs is logical. Two such devices (AutohalerTM (3M Pharmaceuticals, St Paul, MN, USA) and EasibreatheTM (Ivax, Miami, FL, USA)) are currently being marketed and several others are in development [11]. With BA pMDIs, the patient's inhalation through the device triggers a mechanism that fires the pMDI, so that firing and inhaling are automatically coordinated. These devices can achieve good lung deposition and clinical efficacy in patients unable to use a standard "press and breathe" pMDI correctly because of coordination difficulties [23]. BA pMDIs do not solve cold Freon problems and would be unsuitable for a patient who has this kind of difficulty using pMDIs. However, errors when using BA pMDIs are less frequent than when using a standard pMDI [24, 25]. It is essential that the BA pMDI is correctly prepared (e.g. by raising the priming lever, removing the mouthpiece cover etc.); the inhalation must also be strong enough to trigger the firing mechanism (the triggering flow is 2030 L·min1 for currently available devices).
Spacer devices are attachments to the inhaler mouthpiece with a volume ranging from 20750 mL. Their design and performance has been discussed in detail elsewhere [26, 27]. Many have a one-way valve in the mouthpiece, which prevents the patient blowing the dose away after firing. However, inhalation must be strong enough to trigger the one-way valve, otherwise no dose will be delivered. Spacers overcome coordination problems because inhalation can take place either as the device is fired into the spacer or after a short pause, with the latter method being recommended for some models. Cold Freon problems are unlikely with spacer devices because the point of aerosol generation is more remote from the mouth compared with a pMDI. Tidal breathing from the spacer after firing a dose may be acceptable for some models [28] but multiple actuations, long delays between firing and inhaling, and the accumulation of static charge on some plastic spacer devices are likely to reduce the dose available for inhalation [29, 30]. Ideally, each pMDI dose should be inhaled separately from the spacer. Specific handling and washing techniques for different spacers are generally recommended to minimise static charge build-up. While spacers are good drug-delivery devices, they suffer from the obvious disadvantages of making the entire delivery system less portable, compact and convenient than a standard pMDI.
DPIs were first introduced in 1970, and the earliest models were single-dose devices containing the powder formulation in a gelatine capsule, which the patient loaded into the device prior to use. Since the late 1980s, multi-dose devices have been available, giving the same degree of convenience as a pMDI. The first of these was the TurbuhalerTM (AstraZeneca, Lund, Sweden). By early 2005, at least 17 DPIs were marketed in different countries, consisting of both single-dose and multi-dose models [31]. Most multi-dose DPIs hold the powder in a reservoir, from which individual doses are metered, but the Diskus® (GlaxoSmithKline, North Carolina, NC, USA) holds the doses in individually sealed foil blisters. DPIs may be more expensive than pMDIs but this will vary according to pricing policies in different countries. All currently marketed DPIs are breath-actuated and no propellants are needed to generate the aerosol [32]. The patient's inhalation through the device is used to disperse the powder formulation and to deliver it into the lungs. However, patients can make crucial errors with a DPI; for instance, by failing to load a dose correctly or by exhaling into the DPI so that the dose is blown away [33]. Unless clearly instructed, some patients might not know that they must firmly seal their lips around the mouthpiece, causing them to attempt an "open mouth" inhalation technique that will not deliver any dose. The TurbuhalerTM DPI, and possibly other DPIs in which doses are metered from a bulk powder reservoir, must be kept upright (held vertically) when loading a dose before inhalation, so that the dosing chamber will fill under gravity [34]. Compared with a standard pMDI, fewer patients demonstrate errors in inhaler technique with a DPI [24, 25]. Many DPIs must be stored in a dry environment to prevent the drug formulation being degraded by moisture. DPIs tend to work better with rapid and forceful inhalation, since this disperses the powder formulation into small "respirable" particles as efficiently as possible [31]. Delivery to the lungs may be reduced with slow inhalation and for each DPI it is necessary for patients to attain a minimum inhaled flow rate in order to ensure that some drug is delivered to the lungs [6]. It is also desirable that the rate of increase of inhaled flow at the start of inhalation should be as high as possible [35]. This is sometimes called high flow "acceleration" or high "early flow". In theory, the need to inhale forcefully could be a problem for some patients, especially those with more severe obstructive airways disease. However, a recent review [36] concluded that most patients can generate sufficient flows through DPIs to benefit from them, including 98% of asthmatic patients undergoing severe exacerbations needing hospitalisation. Most studies in which patients' inhaled flow rates through DPIs were assessed, have involved asthmatic patients. However, in one study [37], COPD patients with a mean forced expiratory volume in one second (FEV1) of 0.7 L could generate peak inhaled flow rates of 2878 L·min1 via the TurbuhalerTM DPI.
The development of soft mist inhalers (SMIs) has opened up new opportunities for inhaled drug delivery. SMIs use liquid formulations similar to those in nebulisers, but are generally multi-dose devices that have the potential to compete with pMDIs and DPIs in the portable inhaler market. While a number of SMIs are known to be in development [38, 39], the only device currently marketed is Respimat® Soft MistTM Inhaler (Boehringer Ingelheim GmbH & Co. KG, Ingelheim, Germany). This device contains sufficient doses of a bronchodilator formulation for 1 month's dosing, stored in a fluid reservoir [40]. Respimat® Soft MistTM Inhaler is powered by the energy of a compressed spring inside the inhaler; no propellants are required. Individual doses are delivered via a precisely engineered nozzle system as a slow-moving aerosol cloud (hence the term "soft mist"). The velocity of the spray from Respimat® Soft MistTM Inhaler is only about one-tenth of that from a CFC-based pMDI [10]. However, scintigraphical studies have shown that lung deposition is several times higher than that from a CFC-based pMDI [41], and clinical trials have confirmed that drugs delivered by the Respimat® Soft MistTM Inhaler are effective in correspondingly smaller doses in COPD patients [42].
Respimat® Soft MistTM Inhaler is a "press and breathe" device, and the correct inhalation technique closely resembles that used with a pMDI. While coordination between firing and inhaling is required, the low spray velocity and long duration of the aerosol cloud (typically 11.5 s) should enable patients to coordinate firing and inhaling more easily than with a pMDI [43]. Respimat® Soft MistTM Inhaler has been used relatively little in clinical practice to date, and could prove to have advantages or disadvantages additional to those listed in table 2
A variety of inhaler devices are now available to deliver inhaled drugs to patients with COPD. The inhaled drug delivery field is a dynamic one, with many inhalers available already and new ones being introduced on a regular basis. The plethora of inhaler devices available, requiring different inhalation techniques for optimal drug delivery, may confuse patients and healthcare providers alike, a situation described as "device dementia" [44]. That said, a number of actions or steps are common to all types of devices reviewed in this article (table 4
There is no perfect inhaler, and each has advantages and disadvantages, but there is increasing recognition that a successful clinical outcome is determined as much by choice of an appropriate inhaler device as by the drugs that go in them [46]. Drug delivery from all inhaler devices depends on how the patient prepares the device and then inhales from it. The relative difficulties in completing these two steps correctly can be shown on a scale (fig. 1
There is evidence that a patient is most likely to use correctly an inhaler that he or she prefers [25], and each patient's choice of device will be determined by individual perceptions of how its advantages and disadvantages balance out. This decision could be quite different to the judgement of a prescriber or a formulator, who may give more weight to technical points. Choice of an inhaler device should therefore take into account the likelihood that patients will be able to use a particular device correctly, cost-effectiveness, preference and likely compliance.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||